ABSTRACT
Objective
To investigate whether the anteversion angle of acetabular prosthesis can be evaluated on the antero-posterior X ray film of common double hip joint.
Methods
Total 32 patients ( 41 hips ) after total hip arthroplasty were selected , including 18 males and 14 females , aged ( 66.2±4.1 ) years. All patients completed the positive X ray film of both hips and plain CT scan of pelvis after operation. Acetabular anteversion was measured by plain CT scan of pelvis , and measured by Saka and other measurement formulas on X ray film.
Results
The acetabular anteversion measured by X ray film was ( 16.2±5.0 ) °and that measured by CT was ( 31.8±9.7 ) °(P=0.00 ) . In addition , there was a significant linear correlation between X ray film and CT ( Pearson correlation coefficient ( r=0.84 , P=0.00 ) .
Conclusion
CT can accurately measure the acetabular anteversion , but it has obvious disadvantages , such as large radiation , high cost , phantom CT artifact and so on. Although Saka measurement formula can not directly obtain the accurate acetabular anteversion as CT measurement , it has a high correlation with the acetabular antever- sion measured by CT. Therefore , the method proposed in this study can also preliminarily evaluate the acetabular anteversion.